REVISION HISTORY:
Rev 2.0.47 7-10-2024 • Corrected issue with the IQ355 Motion and over capacity status character.
• VRS - Added the IOR/IOW commands for serial reading the DIO inputs and setting the DIO outputs.
• Corrected issue with print tab menu issue with TMP prompts.
Rev 2.0.46 04-05-2022 • Corrected issue with zeroing in a converted unit with a DLC scale that would change the base unit’s deadload incorrectly.
• Fixed calibration counters not incrementing properly with DLC builds.
• Cleaned up DLC diagnostic menu.
• Added Instance-based backdoor password. Contact Technical Support: 1-866-254-8261 or E-mail: tech@cardet.com for instructions.
Rev 2.0.30 • Moved ID storage preset soft key to shift soft key menu due to interfering with shift soft key Diagnostic and soft key send gross.
• Added menus to the 2XX-IP menu so that when Toledo output was selected they can choose between the four possible formats. • Fixed units key giving incorrect conversion. It was always passing lbs. as the starting unit and now passes the current unit.
• Changed default byte order for 2XX-EIP option cards from 4321 to 1234.
• Fixed Send Gross softkey when using SB500 manual mode. Both softkeys can now be used in Normal and ID Storage modes.
• Added code to default Send Gross port to COM0 (0).
Rev 2.0.25 • Added a root password. Note that the root password default is “disabled”.
• Corrected software issues causing the DAC card voltage/current to be inconsistent when the UNITS key is pressed. This was caused by using a DAC max value in the primary unit instead of finding the DAC max value of the secondary unit. The DAC max of the secondary unit is now implemented and is used when the UNITS key is pressed.
• Also corrected zeroing the scale while using the secondary unit.
Rev 2.0.22 • Added ability to print Gross weight with the push of a soft key labeled ‘SEND GRS’.
Rev 2.0.18 Updates – 1/20/2017 • Fixed a loss of deadload during certain scenarios involving power outages.
• Added more descriptive boot-up error messages when there is a CAN bus error.
Rev 2.0.17 Updates – 9/30/2016
• PWC mode of operation now has ability to use PWC1 as a zero monitor.
• Fixed display bug which occurred when ID prompts exceeded 9 characters.
• Added badge reader back to 225-USA build.
• All ICAN BUILDS – fixed bug that allowed weight to be transmitted in certain “no weight” error conditions.
• 225-SNAP – Added radio diagnostics that measure signal strength to each radio (requires 2XX-SNAP rev 1.5.7).
Rev 2.0.16 Updates – 5/25/2016
• Add below zero status “U” to SMA string.
• Fix issues with FINE SPAN and HI RES when adjusting in high weight range (only applies to scales with 3-pt calibration).
• Bug fixes for SST compensation regarding dpp, stability, and saving upon exit of SETUP.
• Add SST-style adjustment to ICAN for hydraulic-ICAN installations.
Rev 2.0.15 Updates – 4/15/2016
• Fixed bug in the storing of serial ports’ “Gross Only” parameter.
Rev 2.0.14 Updates – 2/16/2016
• Sped up read/write operations on the USBA option card.
• Fixed problem when auto assigning an ID, was not checking for ID on flash drive or SDcard on USBA option card.
Rev 2.0.13 Updates – 1/22/2016
• Fixed the toledo long (8530) continuous output data stream checksumming error.
• The Gross weight was not being written to the flash drive record with more than 1 scale and they were not totalized.
Rev 2.0.11 Updates - 10/26/2015
• Fixed problem with site ID for AWID badge reader.
• Added sleep mode for weight display when an SB500 scoreboard is attached. The key sequence is SHIFT-ESC S to turn on/off sleep mode.
Rev 2.0.10 Updates - 8/25/2015
• Add Status for device installed in 2XX-USBA option card.
- SD=SD CARD
- FD=FLASH DRIVE
- KB=KEYBOARD
• Remove duplicated functions from FUNCTION= menu (*NET, *TARE, *UNITS, *ACCUM, *Z, *z).
Rev 2.0.09 Updates - 8/6/2015
• Enable Outputs 5-8 when USBA option card is installed.
• Fix USBA display bug in Setup Menu #3.
• Allow a WinVRS local to send other scales’ weights to remote (besides SCALE 1).
• Add ability to backup 2XX-SNAP calibration to 225 internal memory for later retrieval (e.g. when an option card is replaced).
• Fix bugs in Talley Roll.
Rev 2.0.08 Updates - 4/29/2015
• Add Gross/Net to tally record.
• Bug fixes in continuous output streams.
Rev 2.0.07 Updates - 12/3/2014
• Fixed ANALOG FAIL error when the GUARDIAN scale type is selected.
• Fixed problem where the conversion factor gets corrupted in FINE SPAN and HIRES modes.
Rev 2.0.06 Updates - 11/20/2014
• Added missing fields to ID Storage transaction file.
• Fixed bug in PWC Preset menu for remote indicator.
• Added byte order selection for 2xx-AB and 2xx-EIP option cards.
• Fixed lockup/display issues with various diagnostic selections in the REVIEW/SETUP Menu.
Version 2.0.05 Updates – 10/8/2014
• Slow down printing in PRINT SETUP to allow slower printers to catch up.
• Bug fixes in print setup.
• Added more efficient wait timer.
• Added selection of 2XX-IP option card for printer.
• Prevent beeper from calling scale functions while in setup.
• Removed code disabling interrupts where possible to prevent beeper from being garbled.
• SnapStream bugs fixed:
- Fine span menu displaying error
- Diagnostic menu lockups
- Diagnostic menu display not clearing unneeded info
- Diagnostic menu storing load cell data incorrectly
- 2XX-SNAP remote indicator no long requires a power-cycle to start receiving weight
Version 2.0.04 Updates – 8/21/2014
• Fixed problem with re-printing ticket.
• Verify zeroing from remote indicator and retry (Local/Remote Operation).
• Get weight and reference from ID record for 2nd and 3rd IDs on USBA card (mode= ID Storage or Package Weigher).
• Allow displaying negative weight on remote display for INTL version.
Version 2.0.03 Updates – 6/9/2014
• Modified directory path for file storage on USB flash drive.
• ETR stores Gross or Net weight and mode.
• Fixed display problem when viewing calibration/configuration counters.
• Fixed timing when doing INTL lamp test to display software info.
• iCan limited to 50 sample\sec was 10.
• Clear totalizer over capacity status when all scales under capacity.
Version 2.0.02 Updates – 2/28/2014
• Fixed problem with zeroing and changing units in local/remote operation.
Version 2.0.01 Updates – 1/29/2014
• Added support for 2xx-USBA option card.
• Added Scale ID setup parameter.
Version 2.0.0
• Initial release for re-designed mainboard.
225-INTL-ICAN.hex
1. Supports ICAN
2. Supports analog and serial scales
3. Supports Electronic Talley Roll (ETR)
4. MODE OF OPERATIONS:
a. Normal
b. ID Storage
c. Digital Fill Control
d. Preset Weight Comparator
e. Check Weigher
f. Batcher
g. Package Weigher
5. Serial output formats:
a. SMA
b. SB400
c. TOLEDO (short and long)
d. IQ355
e. NUMERIC WEIGHT
f. SB500
g. SB250/500M
AVR USB Update 225 Main Software Instructions
(AVR-ISP with USB Cable)
For instructions on how to flash the 225 main PC board software, please select Manuals, and then download this PDF:
AVRUSB Update 225 Main Software